Distance vs Object Placement Concept:
The optical resolution (D:S) of our product is 20:1, it refers to the ratio of the measuring distance to the diameter of the measured object. When the distance between the infrared temperature sensor and the object to be measured increases, the surface area of the object to be measured is required to be larger.



4-20mA Model Connection Diagram:

RS485 Model Connection Details:

Slave ID : 1 (can be changed via software or modbus command)
Baud Rate : 4800 (can be changed via software or modbus command)
Register Type : 03 (Coil Register)
Address : 40001, 40002 (Read 2 bytes from 1st register, Temperature in Floating Point)